MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A22040A
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/12/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that at the end of a therapeutic transurethral resection of the prostate (turp) procedure, it was noticed that the ceramic insulation at the distal end of the inner sheath had broken off inside the patient's bladder.However, no fragment remained inside the patient since it was reportedly retrieved with grasping forceps.The intended procedure was successfully completed with the same set of equipment and there was no adverse event or patient injury.In addition, it was reported that the inner sheath had been inspected before the procedure and that no abnormalities or irregularities were detected at that time.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for investigation but to olympus medical systems corporation (omsc), (b)(4) (returned to omsc on 2017-10-17).The evaluation/investigation confirmed that the ceramic insulation at the distal end of the inner sheath had broken off.However, no part was missing since the fragment was returned as well.The cause of this damage and the breakage of the ceramic insulation is most likely excessive high-frequency output power and mechanical overload in combination with improper handling.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the inner sheath without showing any abnormalities.It is clearly stated as a warning note in the instructions that impact, fall, shock or similar stress can damage the ceramic insulation at the sheath's distal end and that the instrument must not be used if damaged since otherwise this can cause injuries to the patient and/or user.The user apparently did not follow these instructions since the damage and breakage of the ceramic insulation were caused by thermal and mechanical overload.Therefore, this event/incident was attributed to use error and the case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes.In addition, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 
4940669662
MDR Report Key6978957
MDR Text Key91296216
Report Number9610773-2017-00122
Device Sequence Number1
Product Code FJL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Catalogue NumberA22040A
Device Lot Number061W
Other Device ID Number04042761029339
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-