Catalog Number 5700000000 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Improper or Incorrect Procedure or Method (2017); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/29/2017 |
Event Type
malfunction
|
Event Description
|
It was reported that the scale was inaccurate due to the shipping pins being left in preventing the litter from resting correctly on the load cells.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Event Description
|
It was reported that the scale was inaccurate due to the shipping pins being left in preventing the litter from resting correctly on the load cells.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|