• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) MAESTRO FOOT SWITCH; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) MAESTRO FOOT SWITCH; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004218500
Device Problem Device Remains Activated (1525)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that the footswitch cannot deactivate rf delivery.The target lesion was located in the floor of the right atrium.A maestro foot switch and a maestro 4000¿ controller were selected for use.During the procedure, a successful typical flutter cti ablation using a maestro 4000 and a foot pedal, along with a mifi nav xp and a rhythmia system was performed.After approximately 30 lesion, physician stepped off the device.However, rf energy continued to deliver.It was stopped by shutting down the generator remotely.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO FOOT SWITCH
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6980419
MDR Text Key90957816
Report Number2134265-2017-10511
Device Sequence Number1
Product Code OAD
UDI-Device Identifier08714729862437
UDI-Public08714729862437
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM004218500
Device Catalogue Number21850
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/09/2017
Initial Date FDA Received10/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age75 YR
-
-