BOSTON SCIENTIFIC - FREMONT (CE) MAESTRO FOOT SWITCH; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004218500 |
Device Problem
Device Remains Activated (1525)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
|
|
Event Description
|
It was reported that the footswitch cannot deactivate rf delivery.The target lesion was located in the floor of the right atrium.A maestro foot switch and a maestro 4000¿ controller were selected for use.During the procedure, a successful typical flutter cti ablation using a maestro 4000 and a foot pedal, along with a mifi nav xp and a rhythmia system was performed.After approximately 30 lesion, physician stepped off the device.However, rf energy continued to deliver.It was stopped by shutting down the generator remotely.No patient complications reported.
|
|