The customer complained of an erroneous high result for 1 patient sample tested for gluc3 glucose hk (gluc3) on a cobas integra 400 plus.The initial gluc3 result was 193.37 mg/dl.This result was reported outside of the laboratory.The patient requested that the sample be repeated.The customer repeated the sample twice with results of 103.88 mg/dl and 103.93 mg/dl.There was no allegation that an adverse event occurred.The gluc3 reagent lot number was 23629001 with an expiration date of 31-jul-2017.On (b)(6) 2017 the customer also had issues with high ldl results; however these results were not discrepant.Calibration and quality controls (qc) were acceptable.During a review of the alarm trace, frequent "no fluid/not enough" alarms occurred.Based on this alarm, it is possible that the sample probe transferred an incorrect amount of sample volume causing a higher level.It is possible the tube settings do not match the filling volume which can cause the sample probe to go too far into the sample.Product labeling states only one type of sample tube or cup can be used in each rack.Different tube types must not be mixed on the same rack.The customer has checked their rack settings and they seem to be correct.The customer uses a clotting time of 20 minutes which was most likely too short.Based on the information available, a general reagent problem can be ruled out since calibration and qc were acceptable.
|