(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: a visual and functional inspection was performed on the returned device.The reported loose or intermittent connection was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during preparation the inflation device was over torqued during connection to the hub resulting in the noted crack on the hub (and subsequent leak) thus resulting in the reported loose or intermittent connection.There is no indication of a product quality issue with respect to design, manufacture or labeling.
|
It was reported that a 3.50 x 18 mm xience alpine stent delivery system (sds) was selected for the procedure.There was no damage noted to the to any of the packaging (chip board box, pouches or dispenser coil) prior to opening and no damage to the hub was observed prior to use.The sds was advanced to the lesion and while attempting to deploy the stent a leak of contrast was observed at the connection between the hub and the indeflator.The sds was removed from the anatomy and replaced with an unspecified sds to successfully complete the procedure.There was no reported clinically significant delay in the procedure.There was no reported adverse patient sequela.No additional information was provided.Returned goods analysis found that when the device was pressurized, fluid leaked out of a crack on the hub.
|