Catalog Number K1023 SMN 10445160 |
Device Problem
Low Test Results (2458)
|
Patient Problem
Missing Value Reason (3192)
|
Event Date 10/07/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted the siemens healthcare diagnostics customer care center about the discordant falsely depressed ca results.The cause of the discordant falsely depressed ca results is unknown.The incident is under investigation by the siemens headquarters support center (hsc).
|
|
Event Description
|
Discordant falsely depressed calcium (ca) results were obtained on patient samples on the dimension vista 1500 instrument.The results were reported to the physician(s) who did not question the results.The same samples were repeated on the same instrument with an alternate ca flex reagent cartridge.Higher results were obtained.Corrected reports were issued.There are no known reports of patient intervention or adverse health consequences due to the discordant falsely depressed ca results.
|
|
Manufacturer Narrative
|
The siemens healthcare diagnostics technical solutions laboratory (tsl) and headquarters support center (hsc) conducted further investigation on lot 17171bd calcium (ca) flex reagent cartridges returned from the customer.Headquarters support center (hsc) has confirmed that when using the dimension vista calcium (ca) reagent lot 17171bd, quality control (qc) and patient samples may show decreased recovery of >0.6 mg/dl (0.15 mmol/l) within an individual well set.The issue does not occur in every flex sequence and not every well set.Siemens is continuing to investigate the reagent issue.
|
|
Manufacturer Narrative
|
Original mdr 2517506-2017-00765 was filed 11/01/2017.Mdr 2517506-2017-00765 supplement 1 was filed 11/30/2017.Siemens healthcare diagnostics has confirmed that dimension vista® calcium (ca) flex® reagent cartridge lot 17171bd may produce erroneously low results from specific well sets.If ca reagent calibration is performed using an unaffected well set and qc and patient samples are subsequently processed using an affected well set, ca results may be falsely depressed up to -2.8 mg/dl (-0.7 mmol/l).If ca reagent calibration is performed using an affected well set, and qc and patient samples are subsequently processed using an unaffected well set, ca results may be falsely elevated.The observed bias for serum, plasma, and urine specimens are similar.Urgent medical device recall, vc-18-03.A.Us and urgent field safety notice vc-18-03.A.Ous dated january 2018 were issued to customers who had been shipped dimension vista® calcium (ca) flex® reagent lot 17171bd.Customers were instructed to discontinue use and to discard lot 17171bd.
|
|
Search Alerts/Recalls
|