A healthcare provider (hcp) reported a patient with an implantable neurostimulator (ins) had a return of their nausea and vomiting about two months prior to the report.When interrogated by the clinician programmer, it indicated the serial number was incorrect, ma was at 0.0, and impedance was showing out of range.The patient had a new device around (b)(6) 2016.The hcp said there was no verbiage showing that there had been a reset of the device.Impedance was tested and showed 601 ohms with recommended current of 3.0ma.They planned to program the patient again and would consult with another hcp.There was no known activity that prompted the event and no procedures or treatments.No further complications were reported/anticipated.
|