• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult To Position (1467); Fitting Problem (2183); Device-Device Incompatibility (2919)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2015
Event Type  malfunction  
Manufacturer Narrative
Zimmer biomet complaint- (b)(4).The parts passed inspection of the outer profile with the exception of some deformed areas where the liner is damaged.This damage likely occurred during the procedure.Complaint sample was evaluated and the reported event was confirmed.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted late as it has been identified in remediation.
 
Event Description
It was reported that during an initial total hip arthroplasty the surgeon was unable to seat the liner in the cup.When he took the liner out he noticed the locking system in the poly had an imperfection.Another liner was opened and used to finish the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 36MM E
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6995832
MDR Text Key91345298
Report Number0001822565-2017-07596
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/05/2019
Device Model NumberN/A
Device Catalogue Number010000935
Device Lot Number3467512
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received02/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/29/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-