The investigation determined higher and lower than expected vitros phyt quality control results were obtained from two different quality control fluids on a vitros 5600 integrated system.The most likely assignable cause is instrument related, as the results of the phyt within-run precision tests were outside of ortho guidelines, indicating the vitros 5600 integrated system was not performing as intended.Ortho field service performed service actions to the immuno-wash subsystem, which included the replacement of the immunowash metering pump and performing all necessary adjustments.Acceptable phyt performance was obtained following these service actions.The investigation found no evidence the vitros phyt reagent malfunctioned.
|
A customer observed higher and lower than expected vitros phyt quality control results obtained from two different levels of non-vitros biorad controls, lot 40912, on a vitros 5600 integrated system.Non-vitros br l2 lot 40912 vitros phyt results 9.9, 8.4, and 9.9 ug/ml versus the expected phyt result 12.58 ug/ml.Non-vitros br l3 lot 40913 vitros phyt results 30.8, 35.2, 29.7, 31.6, 34.3, 30.9, 28.8, 30.9, 34.5, 33.2, 36.6, 30.8, and 33.5 ug/ml versus the expected phyt result 23.48 ug/ml.Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected.Ortho was not made aware of any allegation of patient harm as a result of this event, however, the investigation could not rule out that patient samples were not, or would not be affected if the event were to recur undetected.(b)(4).
|