Catalog Number QD8
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that the bearings were worn.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the attachment device stopped working.It was reported that the device started working after being washed.It was not reported if there was a delay in the procedure due to the event or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CM ANGLE ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6999160
MDR Text Key91449302
Report Number1045834-2017-12374
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberQD8
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2017
Date Manufacturer Received10/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-