Model Number E602
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer received a questionable low elecsys testosterone ii assay result for one patient sample.The initial result for an aliquot of the sample processed by the modular preanalytic analyzer (mpa) was 8 ng/dl and was reported outside of the laboratory.On (b)(6) 2017, the physician questioned the result and the customer repeated the sample.The result from the aliquot of the sample was 624 ng/dl.The result from the primary sample tube was 598 ng/dl.The repeat results were believed to be correct.The patient was not adversely affected.The reagent lot number was 19105300 with an expiration date of 28-feb-2018.The field service representative could not find a cause.He ran mechanical checks and performance testing which passed, checked the probe positions, and primed the system.
 
Manufacturer Narrative
A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.Typical causes for this type of event include issues with sample quality or insufficient maintenance of the analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 E 602 MODULE
Type of Device
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6999860
MDR Text Key92073855
Report Number1823260-2017-02526
Device Sequence Number0
Product Code CDZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Medical Technologist
Type of Report Initial,Followup
Report Date 11/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberE602
Device Catalogue Number05990378001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Age69 YR
-
-