• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER Back to Search Results
Device Problem Insufficient Information (3190)
Patient Problem Burn(s) (1757)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
The employee subject of the reported event sought and received medical treatment and returned to work.The employee was not wearing the required ppe at the time of the reported event.The user facility stated that the v-pro max sterilizer was operating properly.Th e steris operator manual states on page 1-2, "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.Refer to the vaprox hc sterilant safety data sheet (sds) for appropriate personal protective equipment (ppe; refer to section 2, terms, definitions and symbols), spill containment and cleanup." and on page 6-14, "steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical-resistant gloves when using the sterilization unit".A steris account manager notified the user facility of the appropriate ppe required when using the v-pro max sterilizer.No additional issues have been reported.
 
Event Description
The user facility reported an employee experienced a burn when unloading their v-pro max sterilizer.No procedure delay or cancellation was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7002705
MDR Text Key92478174
Report Number3005899764-2017-00058
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/18/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-