Model Number 0010-2090
Device Problems Material Perforation (2205); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2017
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: failure analysis for fiber (b)(4): the fiber cap exhibits signs of melting which formed a hole from the surface to the bevel edge; the glass cap exhibits severe devitrification at the output area, and detritus adhesion around output area; the heat shrink tubing exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
 
Event Description
It was reported that during a bph surgical procedure "premature failure" was noted.The fiber was exchanged and two additional "premature failures" were reported.A fourth fiber was used to complete the procedure.Patient outcome: "no injury" reported.This report is for the second fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT HPS BPH FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
jennifer mascioli-tudor
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7009536
MDR Text Key92315621
Report Number2937094-2017-00966
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953004709
UDI-Public00878953004709
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/23/2018
Device Model Number0010-2090
Device Catalogue Number0010-2090
Device Lot Number648H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2017
Date Manufacturer Received10/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-