Model Number 28-95-85 |
Device Problems
False Alarm (1013); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 control panel mrp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that an s5 control panel mrp displayed an error and an alarm sounded during a procedure.There was no tubing in the pump and the device was not in operation.The customer reset the occlusion and the device began to function as expected.There was no report of patient injury.
|
|
Manufacturer Narrative
|
No investigation based on the readout of the pump could be performed due to the age of the device and its components: this pump had the old version of the processor board and with this version it is only possible to store one error per day.As the issue could not be identified in the readout of the system, a root cause was not identified.A review of the dhr did not identify any deviations or nonconformities relevant to the issue.
|
|