It was reported that the generator was not detecting heart rate.The internal data of the generator was reviewed and it was determined that low impedance had been detected on the device in the past as well as a large drop in impedance.Low impedance may interfere with heartbeat detection.The generator was later able to detect heart rate when the sensitivity was set to its least sensitive detection setting (sda sensitivity - 1).However, approximately four months later, the generator continued to have difficulty detecting heart rate.Low impedance was detected intermittently through system diagnostics.The patient reported that approximately a year ago she had felt a 'movement' in her neck, where the lead connected to the nerve and when she moved her head back, the movement reversed and 'went back.' the patient said that she still felt normal and magnet mode stimulation.No known relevant surgical intervention has occurred to date.No further relevant information has been received to date.
|