PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number GL358M
Device Problem Fire (1245)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2017
Event Type  malfunction  
Manufacturer Narrative
The device has not been made available for evaluation at this time.Should further information or the device become available, a follow-up report will then be submitted.
 
Event Description
Alleges transformer caught on fire and melted on the end-user's hardwood floor also causing burn marks to it.
 
Manufacturer Narrative
The device was returned and evaluated.The evaluation concluded that the transformer was subjected to liquid ingress.
 
Event Description
Alleges transformer caught on fire and melted on the enduser's hardwood floor also causing burn marks to it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave.
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a,
Manufacturer Contact
kelly livingston
401 york ave
n/a
duryea, PA 18642
8008008586
MDR Report Key7013023
MDR Text Key91430379
Report Number2530130-2017-00143
Device Sequence Number1
Product Code INO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070950
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberGL358M
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-