Catalog Number BI70002000 |
Device Problems
Device Inoperable (1663); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information not provided due to (b)(6) patient privacy regulations.The site called the field representative and they were able to resolve the issue over the phone by changing the ip address to the correct one.The site reported that a system reboot following that resolved the issue.No further issues were reported.
|
|
Event Description
|
A medtronic representative reported that while in a procedure, both the navigation system and the imaging system became unresponsive.There was no impact on patient outcome.There was a reported delay to the procedure of less than 1 hour due to this issue.
|
|
Manufacturer Narrative
|
Additional procedure information received.
|
|
Event Description
|
Medtronic received information that the reported issue occurred during a lumbar posterior instrumented fusion procedure.The navigation and imaging systems became unresponsive while the imaging system was acquiring a spin.Rebooting both systems caused a 10 minute delay.
|
|