The company service representative examined the system and found system message (sm) - ¿lamp calibration data is corrupted: lamp needs to be calibrated.Please contact field service¿ displayed in the event log.The company service representative restarted the system and the sm did not display.However, the company service representative replaced the illuminator module as a preventative measure.The system was then tested and met all product specifications.The system was manufactured on november 23, 2009.Based on qa assessment, the product met specifications at the time of release.The illuminator module was received and a visual assessment of the returned sample revealed no obvious nonconformity.The sample was installed into a calibrated system for functional testing.Upon boot-up the system displayed sm ¿lamp calibration data is corrupted: lamp needs to be calibrated.Please contact field service¿; however, this sm is only an advisory and displays normally when a new illuminator is installed into the system and thus does not indicate an issue.The system was rebooted but the sm did not display again.The module was then tested and found to meet specifications.The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|