A nurse reported the intraocular pressure was not being maintained due to an unspecified infusion issue during a vitrectomy procedure.The product was replaced, the intraocular pressure was maintained and the procedure was completed.There was no harm to the patient.Additional information was requested; however, none has been received to date.
|
The lot complaint history was reviewed, this is the fifth complaint for the finish goods lot; however, the second for this issue.The device history records shows the product was released per specifications.The used returned sample was visually inspected and surgical residue and balance salted solution (bss) crystals were in the fluid path of the manifolds.The filters in the cassette were saturated with bss.It was noted the infusion cannula was not returned.Full functional and performance testing of the sample was then performed on a console.The ball in the check valve of the drip chamber moved freely per specification.The sample could prime and pass intraocular (iop) calibration successfully.No anomalies were observed during priming.The infusion pressure was measured at multiple set points throughout the console range and met specifications.The value of 7500cpm cut rate, 650mmhg wetant, 650mmhg extrusion, and, 650mmhg vacuum, 650mmhg continuous reflux, and 120mmhg proportional reflux displayed on the progress bar.Toggling the infusion and the fluid/air exchange (f/ax) modes, fluid and air flowed from the cassette to the infusion line continuously without any bubble in various settings in all sub modes.The iop was stable in 15 - 16 cc/min at 30 mmhg setting during f/ax testing.No message code appeared on the screen.Fluid flowed from bss to the drain bag without any interfering.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample met specifications.After a thorough investigation of this complaint, it has been determined that this sample met specifications; therefore, no corrective action is required at this time.The manufacturer internal reference number is: (b)(4).
|