PENUMBRA, INC. POD PACKING COIL; HCG, KRD Back to Search Results
Catalog Number RBYPODJ15
Device Problems Mechanical Problem (1384); Device Dislodged or Dislocated (2923); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/23/2017
Event Type  malfunction  
Manufacturer Narrative
The product was disposed of by the hospital and is no longer available for return.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
 
Event Description
The patient was undergoing a coil embolization procedure for a vessel sacrifice in the gastroduodenal artery (gda) using a pod packing coil (podj).During the procedure, while attempting to place the podj behind a ruby coil using a lantern delivery microcatheter (lantern), the podj ended up kicking the lantern out of the target vessel; therefore, it was removed.The procedure was completed using a ruby coil and the same lantern.It should be noted that the physician used a rotating hemostasis valve (rhv) and used a continuous flush throughout the procedure.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7021012
MDR Text Key92730339
Report Number3005168196-2017-02020
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548017648
UDI-Public00814548017648
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberRBYPODJ15
Device Lot NumberF79420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2017
Initial Date FDA Received11/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
-
-