Catalog Number RBYPODJ15 |
Device Problems
Mechanical Problem (1384); Device Dislodged or Dislocated (2923); Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was disposed of by the hospital and is no longer available for return.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure for a vessel sacrifice in the gastroduodenal artery (gda) using a pod packing coil (podj).During the procedure, while attempting to place the podj behind a ruby coil using a lantern delivery microcatheter (lantern), the podj ended up kicking the lantern out of the target vessel; therefore, it was removed.The procedure was completed using a ruby coil and the same lantern.It should be noted that the physician used a rotating hemostasis valve (rhv) and used a continuous flush throughout the procedure.There was no report of an adverse effect to the patient.
|
|