KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-092-12-91
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
The head of the mmf screw broke off near the top 2 threads during insertion.Base of the screw was removed from the patient.
 
Manufacturer Narrative
Product was returned.An investigation was performed using visual and stereo microscopic inspection on the product returned.Further investigation determined that there was no indication of material or method and results of investigation manufacturing defects observed.The device lot number was not provided by the customer therefore the device history records could not be reviewed.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Type of Device
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM   78570
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key7021827
MDR Text Key91786539
Report Number9610905-2017-00086
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00888118037220
UDI-Public(01)00888118037220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number25-092-12-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/13/2017
Event Location Hospital
Date Report to Manufacturer10/13/2017
Date Manufacturer Received11/23/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age27 YR
-
-