Model Number C12058
Device Problems Fracture (1260); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Product investigation: the results of the investigation concluded that the tip coil of the radiopaque tip had been fractured and separated from the distal of the jacket; the fractured tip coil was not returned.The corewire had been subsequently fractured.The distal portion of the fractured corewire was not returned.The distal corewire and shaft had been bent at multiple locations.Additional investigation revealed that the wire was exposed to excessive torque and manipulation and there is evidence of welding on the proximal joint between the coil and the corewire.Torqueing or excessive manipulation of the guidewire in a sharp bend, against resistance, or repeated attempts to cross a total vessel occlusion may damage and/or fracture the pressurewire, which is inconsistent with the instructions for use (ifu).The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the guidewire damage is consistent with forcible contact during use.The pressurewire aeris instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire aeris instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire aeris instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressurewire if it meets resistance.
 
Event Description
The pressurewire aeris became caught in the stent strut.There was difficulty removing the device from the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7025334
MDR Text Key92752493
Report Number3008452825-2017-00288
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberC12058
Device Catalogue NumberC12058
Device Lot Number5765083
Other Device ID Number0541473405577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-