• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. RONGEUR PITUITARY CER BLACK SERRATED 3MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. RONGEUR PITUITARY CER BLACK SERRATED 3MM Back to Search Results
Catalog Number 8733-7183
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that the jaw of a rongeur was found to be misaligned after cleaning and sterilization.There were no surgical or patient impacts associated with this event.
 
Manufacturer Narrative
Additional information: lot number; the returned rongeur was functionally and visually evaluated.When the handles were actuated, the jaw opened and closed as expected.However, the upper jaw was found to be slightly bent so the teeth did not align.The cause is likely attributed to damage from use or handling.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RONGEUR PITUITARY CER BLACK SERRATED 3MM
Type of Device
RONGEUR
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7026322
MDR Text Key92856649
Report Number3012447612-2017-00654
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number8733-7183
Device Lot NumberIM00077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-