Catalog Number 8733-7183 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
|
Event Description
|
It was reported that the jaw of a rongeur was found to be misaligned after cleaning and sterilization.There were no surgical or patient impacts associated with this event.
|
|
Manufacturer Narrative
|
Additional information: lot number; the returned rongeur was functionally and visually evaluated.When the handles were actuated, the jaw opened and closed as expected.However, the upper jaw was found to be slightly bent so the teeth did not align.The cause is likely attributed to damage from use or handling.A review of the manufacturing records did not identify any issues which would have contributed to this event.
|
|