The customer contacted the siemens customer care center (ccc).The customer's quality control (qc) was out low of the acceptable limits for level 1 and level 2 when the event occurred.A siemens customer service engineer was dispatched to the customer's site.The cse installed clean repair kit and power flushed the integrated multi-sensor technology (imt) system.The cse ran dilution check several times and rebuilt imt rotary valve.The cse inspected the imt mixer.The cse replaced the imt pump tubing and aligned it.The cse performed another power flush with hot water, checked the rotary valve after which the leak test passed.The cse ran precision, resulting poor with the first result low.The cse then replaced the rotary valve, after which sodium (na) was running high.The cse replaced the imt module and ran qc, resulting within specification.The cse auto aligned the imt module.The cse primed the imt fluids, ran calibration and dilution check, resulting satisfactory.The cse ran qc, resulting within range.The cause of the discordant, na results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|