Product analysis: upon receipt at medtronic¿s quality laboratory, the valve did not show evidence of impingement that may have contributed to leaflet motion.The carbon assembly, as received, did not rotate in the sewing ring.Green and white multifilament sutures were attached on the sewing ring.Both leaflets were received in the closed position.Using a blue actuator to test leaflet movement, the leaflets moved without difficulty.Both leaflets were intact with no evidence of damage or surface anomalies.Both inflow and outflow valve hinge mechanisms were intact.The inflow rim was intact.The outflow rim exhibited a fracture.The sewing cloth and lock ring area adjacent to the rim fracture showed damage.The lock ring was pulled away from the carbon assembly.The lock ring also showed teeth marks consistent with damage caused by an instrument/tool such as a hemostat clamp during explant.The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Based on the analysis, the clinical observation could not be confirmed; no evidence of leaflet motion issue was noted.With the information available, the reported leaflet motion issue could be due to sizing.If information is provided in the future, a supplemental report will be issued.
|