BOSTON SCIENTIFIC - MAPLE GROVE NC EMERGE®; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493927608350
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event : 18 years or older.(b)(4).The device was not received for analysis.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that balloon pinhole occurred.The target lesion was located in the left anterior descending (lad) artery.A 3.50mm x 8mm emerge balloon catheter was advanced for dilation; however, upon inflation at 12 atmospheres, the balloon did not expand.The physician then concluded that there was a pinhole on the balloon.The device was completely removed and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE®
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7029529
MDR Text Key92517546
Report Number2134265-2017-11392
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Model NumberH7493927608350
Device Catalogue Number39276-0835
Device Lot Number20498030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-