Catalog Number A103200000000
Device Problems Incorrect Measurement (1383); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply (b)(4).The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a xsmart plus goes into auto reverse mode directly once the file inserted in the canal; no injury resulted.
 
Manufacturer Narrative
Evaluation found a broken casing/housing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART PLUS
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
Manufacturer (Section G)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ   1338
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7029617
MDR Text Key93164819
Report Number8031010-2017-00118
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeIZ
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA103200000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-