Catalog Number 9735018
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information not provided due to (b)(6) patient privacy regulations.The suspect suction has been received by manufacturer for evaluation.When the received instrument was connected to a good known system, the suction was recognized but would not track returning a red status.A check of the em interface revealed a issue with coil causing the poor signal.
 
Event Description
A medtronic representative reported that during transnasal endoscopic pituitary surgery, the malleable suction was successfully registered at the beginning of the procedure however during the case the suction had recognition failure.A different malleable suction was used to complete the procedure.There was no delay to procedure.No impact on patient outcome.
 
Manufacturer Narrative
The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLEABLE SUCTION MEDIUM, BALL TIP
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7031200
MDR Text Key92873649
Report Number1723170-2017-04653
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169443532
UDI-Public00643169443532
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133665
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735018
Device Lot Number160509
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-