STRYKER ENDOSCOPY-SAN JOSE PKG., ASSY., STRYKEFLOW 2 WITH TIP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Catalog Number 0250070520
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there was smoking at the handpiece around the buttons.The procedure was completed successfully without any adverse consequences.
 
Manufacturer Narrative
Alleged failure: used during case.Surgitech heard noise and then smoking at handle around buttons.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause could be fluid ingress.The product was returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.(b)(4).
 
Event Description
It was reported that there was smoking at the handpiece around the buttons.The procedure was completed successfully without any adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., ASSY., STRYKEFLOW 2 WITH TIP
Type of Device
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
valerie estrada
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7031620
MDR Text Key92724178
Report Number0002936485-2017-01103
Device Sequence Number1
Product Code GCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K954726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0250070520
Device Lot Number17196FG2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-