DENTSPLY MAILLEFER LENTULO RA 25MM 3; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A002222500300
Device Problems Break (1069); Material Separation (1562)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that one lentulo ra 21 mm 3 separated; no patient injury reported.
 
Manufacturer Narrative
Involved lentulo is actually broken at the base of the active part.No material defect was found during analysis of the rupture pattern.Nothing unusual to report was found during dhr review.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENTULO RA 25MM 3
Type of Device
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
Manufacturer (Section G)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ   1338
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7032835
MDR Text Key93218914
Report Number8031010-2017-00121
Device Sequence Number1
Product Code EKS
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA002222500300
Device Lot Number1391661
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-