CODMAN AND SHURTLEFF, INC NBCA LIQUID EMBOLIC SYSTEM 2; TISSUE ADHESIVE FOR USE IN EMBOLIZATION OF BRAIN ARTERIOVENOUS MALFORMATIONS
|
Back to Search Results |
|
Catalog Number 631500 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/25/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was returned for analysis; however, the analysis has not yet been completed.Additional information will be submitted within 30 days of receipt.This is 1 of 4 mdrs submitted for this complaint with associated report numbers of 1226348-2017-00194, 1226348-2017-00193, 1226348-2017-00192 and 1226348-2017-00191.
|
|
Event Description
|
As reported by a physician, when injecting glue with tantalum and aprox 1/2 cc to 1cc of oil, the glue was not visible as injecting under fluro.Product included trufill nbca 1 gram kit (631500/ ha2573 x 2, 631500/ hf2890 x 1, and unknown catalog/lot x 1).The embolization procedure was for treatment of vein of galin on a (b)(6) infant.The physician, who has been successfully been using nbca glue for many years, added the entire packet of tantalum powder with oil (approx.½ cc to 1cc of oil) to the nbca glue in a glass beaker.The radiopacity of the mixture was not confirmed via flouro prior to injection by comparing the glue to a similar syringe containing contrast as instructed in the ifu.Four separate injections were used and they appeared very faint via flouro, and in one case, the tantalum appeared to be beaded.It was reported that the oil had been protected from light and the mixture was prepared as per the instructions for use (ifu).Fortunately, the patient was successfully treated with no injury.
|
|
Manufacturer Narrative
|
Four glass beakers, one syringe and four plastic containers were received inside of a plastic bag.They were inspected and they were found with residues powder and dry glue.The rest of the components were not returned for evaluation.The four glass beakers, syringe and the four plastic containers were inspected under microscope and residues of powder gray/glue dried can be observed inside of them.An x-ray was performed on received syringe and the 4 beakers.Residues of the mix of the nbca can be observed.The dhr review of the manufacturing documentation associated with this lot ha2573 presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Additional information will be submitted within 30 days of receipt.
|
|
Manufacturer Narrative
|
Information received on 2/28/2018 confirmed that procedure images could not be obtained.As reported by a physician, when injecting glue with tantalum and approximately 1/2 cc to 1cc of oil, the glue was not visible as injecting under fluoroscopy.Product included trufill nbca 1 gram kit (631500/ ha2573 x 2, 631500/ hf2890 x 1, and unknown catalog/lot x 1).The embolization procedure was for treatment of vein of galin on a (b)(6) infant.The physician, who has been successfully been using nbca glue for many years, added the entire packet of tantalum powder with oil (approx.½ cc to 1cc of oil) to the nbca glue in a glass beaker.The radiopacity of the mixture was not confirmed via fluoro prior to injection by comparing the glue to a similar syringe containing contrast as instructed in the ifu.Four separate injections were used and they appeared very faint via fluoro, and in one case, the tantalum appeared to be beaded.It was reported that the oil had been protected from light and the mixture was prepared as per the instructions for use (ifu).Fortunately, the patient was successfully treated with no injury.Multiple attempts to obtain procedure films to confirm the event were unsuccessful.Four glass beakers, one syringe and four plastic containers were received inside of a plastic bag, they were inspected and were found with residues of powder and dry glue.The rest of the components were not returned for evaluation.The four glass beakers, syringe and the four plastic containers were inspected under microscope and residues of powder gray/glue dried coul be observed inside of them.An x-ray was performed on the received syringe and the 4 beakers.Residues of the mix of the nbca could be observed.The dhr review of the manufacturing documentation associated with the lots presented no issues during the manufacturing or inspection process that can be related to the reported complaint.The expiration date for the subassembly components ethiodized oil 16lu602a and tantalum powder h66202 were reviewed and no anomalies were noted.The nbca lack of radiopacity could not be confirmed since during x-ray of the returned materials, residues of the nbca mixture were radiopaque.Additional testing to confirm adherence to specification for the same lot of nbca glue was completed, and the material met all specifications.The cause of the event could not be conclusively determined.The instructions for use warns: ¿prior to injection it is essential to determine, via fluoroscopy, the radiopacity of the mixture by comparison with a similar syringe containing contrast.Inadequate visualization of the mixture could cause inappropriate embolization.¿ a complaint handling database review found no other similar issues for the past 3 years.There is no current safety signal identified related to the reported event based on review of complaint history for the device.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is 1 of 4 mdrs submitted for this complaint with associated report numbers of 1226348-2017-00194, 1226348-2017-00193, and 1226348-2017-00191.
|
|
Search Alerts/Recalls
|
|
|