• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; BARDEX FOLEY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; BARDEX FOLEY Back to Search Results
Model Number 0165SI16
Device Problem No Flow (2991)
Patient Problem Urinary Retention (2119)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the catheter would not drain urine.The complainant alleged that patient's bladder was scanned and still showed full.Allegedly once catheter was removed and replaced, the patient was able to void.
 
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use state the following: "indications for use: the bardex i.C.Latex foley catheter is indicated for use in the drainage and/or collection and/or measurement of urine.Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as nephrostomy tract.Directions for use: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Using proper aseptic methods, remove catheter from package.Prepare patient per hospital/nursing recommended procedure.Proceed with catheterization using standard techniques.Inflate the balloon with sterile water to the volume stated on the inflation lumen of the catheter by using a water-filled luer tip syringe.Connect catheter to collection container.To deflate the balloon, gently reinsert the luer tip syringe into the valve and aspirate.Caution: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure in the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and /or lead to injury, illness or death of the patient.Valve type: use luer slip tip syringe.Do not use needle.Do not exceed recommended capacities as stated on the inflation lumen of the catheter.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: do not aspirate urine through drainage funnel.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Do not use if package is damaged.Recommended indwelling time not to exceed 28 days." (b)(4).
 
Event Description
It was reported that the catheter would not drain urine.The complainant alleged that patient's bladder was scanned and still showed full.Allegedly once the catheter was removed and replaced, the patient was able to void.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
BARDEX FOLEY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7036630
MDR Text Key92508932
Report Number1018233-2017-05929
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016752
UDI-Public(01)00801741016752
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0165SI16
Device Catalogue Number0165SI16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-