• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL; SEE H10)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL; SEE H10) Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 10/21/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No procode, common device name, unique device identification(udi) and/or 510k provided as this device is not released for distribution in the united states.No parts have been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure, the vertek arm handle was loose but the bowl joint was not functioning normally.There was no patient present at the time of the issue.
 
Manufacturer Narrative
The vertek arm was returned to the manufacturer for evaluation.Testing found that the starburst end of the arm would not tighten when prompted by the user.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CART 9734056 S7 STAFF SHRT 100-120V INTL
Type of Device
SEE H10)
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7039570
MDR Text Key93088785
Report Number1723170-2017-04723
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10)
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 01/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9734056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-