• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL L 10MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number 544995
Device Problems Failure to Align (2522); Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation at this time.Teleflex will continue to monitor and trend related events.
 
Event Description
Issue reported: the jaws are not able to hold the clips, might be because of misalignment issue.There was no patient injury.
 
Manufacturer Narrative
Qn#: (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was manufactured at the (b)(4) facility as part of a (b)(4) pc.Lot in january of 2015.The returned instrument was evaluated and found that the knob and tube assembly move freely back and forth when the handle is squeezed indicating the snap ring which holds the knob & tube assembly to the handle has been popped out of its groove thus validating the complaint.Parts were 100% visually inspected and tested at the (b)(4) facility before instruments were sent to customer.No irregularities were found and or reported at the time of inspection and assembly of the product, as this is a standardized process for all instruments manufactured at this facility.We are unable to determine how this instrument has been stored, handled and cleaned by the end users facility.At this time it is undetermined as to what caused the snap ring to pop and the shaft and knob assembly to become loose from the handle but mishandling at the customers site is suspected.Product failure was caused by failure of a single component, and is an isolated event.Manufacturing other remarks: process of product, or design are not suspected to be at fault.
 
Event Description
Issue reported: the jaws are not able to hold the clips, might be because of misalignment issue.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL L 10MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key7044564
MDR Text Key93084185
Report Number3011137372-2017-00368
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544995
Device Lot Number06A1496937
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-