Qn# (b)(4).The dhr for the alleged instrument was reviewed and found completely without any irregularities.This instrument was manufactured at the teleflex medical, kenosha facility as part of a 50 pc.Lot in october of 2015.We are unable to validate the alleged complaint at this time.The returned instrument was evaluated and found that the jaws are aligned both in the open and closed position and the open jaw gap was measured at (.379) which is within print specifications of (.385 +/-.025)and the closed jaw gap was measured at (.008) which is in print specifications of (.004/.012).Further evaluation of the jaws did not show any signs of damage to either jaw and that this instrument is able to pick-up, retain, close and release multiple clips both with and without the use of silastic tubing.We are unable to replicate the alleged issue therefore we are unable to validate this complaint.Parts were 100% visually inspected and tested at the tecomet, inc.Kenosha facility before instruments were sent to customer.No irregularities were found and or reported at the time of inspection and assembly of the product, as this is a standardized process for all instruments manufactured at this facility.No further action required at this time.
|