STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
|
Back to Search Results |
|
Model Number 174301 |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/20/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to stryker sustainability solutions for evaluation.Visual inspection of the device revealed signs of clinical use.Debris is present on the blades of the device.The device was inspected under magnification and the blades revealed wear.Functional inspection revealed the blades to be dull and unable to provide a clean cut.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: user attempts to cut staples, clips, or non-soft tissue; wear to blades due to clinical use.The instructions for use (ifu) state: do not directly apply current to staples or clips.Instruments were designed for cutting soft tissue.Attempting to cut staples or clips may damage the instrument.If cutting of staples or clips is attempted, damage to instrument may occur.The reported event will continue to be monitored through post-market surveillance.
|
|
Event Description
|
It was reported the laparoscopic scissors were dull.The device was replaced.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|
|
Search Alerts/Recalls
|
|
|