MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number 174301
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/20/2017
Event Type  malfunction  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection of the device revealed signs of clinical use.Debris is present on the blades of the device.The device was inspected under magnification and the blades revealed wear.Functional inspection revealed the blades to be dull and unable to provide a clean cut.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: user attempts to cut staples, clips, or non-soft tissue; wear to blades due to clinical use.The instructions for use (ifu) state: do not directly apply current to staples or clips.Instruments were designed for cutting soft tissue.Attempting to cut staples or clips may damage the instrument.If cutting of staples or clips is attempted, damage to instrument may occur.The reported event will continue to be monitored through post-market surveillance.
 
Event Description
It was reported the laparoscopic scissors were dull.The device was replaced.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
marissa richmond
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key7044911
MDR Text Key93233597
Report Number0001056128-2017-00138
Device Sequence Number1
Product Code NUJ
UDI-Device Identifier00885825012564
UDI-Public(01)00885825012564(17)190721(10)4746828
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/21/2019
Device Model Number174301
Device Catalogue Number174301RR
Device Lot Number4746828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-