Model Number 778726 |
Device Problems
Break (1069); Detachment Of Device Component (1104)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that during the procedure, the stent broke off at the end.There was no reported impact to the procedure or patient.
|
|
Event Description
|
It was reported that during the procedure, the stent broke off at the end.There was no reported impact to the procedure or patient.
|
|
Manufacturer Narrative
|
Received 1 ureteral stent sheath only in open packaging.The reported event was unconfirmed since the stent was not returned with the sample (incomplete device return).During the visual inspection no obvious defects were noted in the sample.The stent was not evaluated visually and dimensionally because it was not received with the sample.The push catheter dimensions were found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "¿ improper handling technique can seriously weaken the stent.Acute bending or overstressing during placement could result in subsequent separation of the stent at the point of stress after a prolonged indwelling period.¿ exercise care.Tearing of the stent can be caused by sharp instruments.¿ care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation.Section" (b)(4).
|
|
Search Alerts/Recalls
|