• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® CRITICORE® MONITOR; CRITICORE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® CRITICORE® MONITOR; CRITICORE MONITOR Back to Search Results
Model Number 000002N
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the screen on the monitor was frozen.
 
Manufacturer Narrative
The reported issue of the screen was frozen on the monitor was unconfirmed.The root cause of the reported issue could not be determined as the unit was inoperable at the time of the initial investigation.The reported issue was further investigated after the coin cell was replaced and the unit powered on.The unit was then operating normally.It was likely that the screen froze as a result of the coin cell voltage being below the 3.00vdc minimum operating voltage, although this could not be confirmed.Operation below the 3.00vdc minimum cannot be guaranteed.The unit has been serviced and is working properly.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "warning: the criticore monitor requires special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following tables.Portable and mobile rf communications equipment may affect the criticore monitor.Warning: the criticore monitor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the criticore monitor should be observed to verify normal operation in the configuration in which it will be used.Warning: do not immerse or submerge the monitor or turn it upside down when cleaning.Caution: use only heavy-duty alkaline d cell batteries.Do not use rechargeable batteries.Do not incinerate batteries.Recycle or dispose of them properly.Contact bard for disposal information.Caution: improper orientation of the batteries within the battery pack can potentially damage the criticore monitor.Caution: state and federal regulations govern the packaging necessary for return of medical product which may have been contaminated.Refer to local, state and federal regulations when packaging the criticore monitor for return.Caution: there are no user serviceable components inside the criticore monitor.The user should not attempt to repair the criticore monitor.To do so may void the warranty and could result in erroneous monitor readings.Caution: use of cables or sensors other than those specified for use with the criticore monitor, except those sold by bard for use as replacement part or repair components, may result in increased emissions or decreased immunity of the criticore monitor.Caution: the criticore monitor should be recycled properly per european union directive 2002/96/ec on waste electronic and electrical equipment, january 27, 2003.Do not dispose with ordinary municipal waste." (b)(4).
 
Event Description
It was reported that the screen on the monitor was frozen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® CRITICORE® MONITOR
Type of Device
CRITICORE MONITOR
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7045668
MDR Text Key93065554
Report Number1018233-2017-06011
Device Sequence Number1
Product Code EXS
UDI-Device Identifier00801741080777
UDI-Public(01)00801741080777
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number000002N
Device Catalogue Number000002N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-