Catalog Number BI70002000 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that viewing station computer was replaced to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer has not been received by the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that while outside of procedure, the mobile view station (mvs) was waiting to communicate with image acquisition system (ias) because the viewing station computer was power cycling.There was no patient present at the time of the issue.
|
|
Manufacturer Narrative
|
The computer for the viewing station of the imaging system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|