Catalog Number 1120250-23
Device Problems Device Dislodged or Dislocated (2923); Physical Property Issue (3008)
Patient Problem No Patient Involvement (2645)
Event Date 11/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that when the protective sheath was removed from the 2.5 x 23 mm xience alpine stent delivery system it felt stiff and pulled the stent off the balloon with it.There was no resistance noted during removal of the protective sheath.The device could not be used on the patient.A new stent delivery system was used to complete the case.There was no clinically significant delay in the procedure reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The reported stent dislodgement was able to be confirmed.The reported physical property issue was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that inadvertent mishandling resulted in the noted bent protective sheath thus as the sheath was being removed resulted in the reported physical property issue/irregular texture.Manipulation to remove the sheath resulted in the noted stent damages (bent, crushed, flared) and ultimately resulted in the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7052678
MDR Text Key93199548
Report Number2024168-2017-09148
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2020
Device Catalogue Number1120250-23
Device Lot Number7090541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-