• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BERCHTOLD GMBH & CO. KG EXAMINATION LIGHT F 300, CEILING MOUNT, LED DIMMABLE, DOUBLE-ARM, 24V DC, 100-; DEVICE, MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BERCHTOLD GMBH & CO. KG EXAMINATION LIGHT F 300, CEILING MOUNT, LED DIMMABLE, DOUBLE-ARM, 24V DC, 100-; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Catalog Number CE 2863075
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2017
Event Type  malfunction  
Manufacturer Narrative
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.The stryker field service technician (sfst) stated that he was provided photos and a statement from the customer that the spring arm disengaged from the horizontal arm.The investigation is ongoing.Once the investigation is complete, a supplemental will be filed.There was no injury or adverse consequence reported.
 
Event Description
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.There was no injury or adverse consequence reported.
 
Manufacturer Narrative
It was reported that an f300 light separated at the spring arm to horizontal arm joint.The connection point was damaged.While it is unknown how the damage occurred, potential root causes could be maintenance of the system by non-stryker personnel (for example, the p3 maintenance company employed by the hospital).The light has been replaced at the site by a stryker field service technician.
 
Event Description
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.There was no injury or adverse consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXAMINATION LIGHT F 300, CEILING MOUNT, LED DIMMABLE, DOUBLE-ARM, 24V DC, 100-
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
Manufacturer (Section G)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
Manufacturer Contact
volker hornscheidt
ludwigstaler strasse 25
tuttlingen 78532
74611810
MDR Report Key7054443
MDR Text Key93110396
Report Number0008010153-2017-00037
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCE 2863075
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-