BERCHTOLD GMBH & CO. KG EXAMINATION LIGHT F 300, CEILING MOUNT, LED DIMMABLE, DOUBLE-ARM, 24V DC, 100-; DEVICE, MEDICAL EXAMINATION, AC POWERED
|
Back to Search Results |
|
Catalog Number CE 2863075 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/31/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.The stryker field service technician (sfst) stated that he was provided photos and a statement from the customer that the spring arm disengaged from the horizontal arm.The investigation is ongoing.Once the investigation is complete, a supplemental will be filed.There was no injury or adverse consequence reported.
|
|
Event Description
|
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.There was no injury or adverse consequence reported.
|
|
Manufacturer Narrative
|
It was reported that an f300 light separated at the spring arm to horizontal arm joint.The connection point was damaged.While it is unknown how the damage occurred, potential root causes could be maintenance of the system by non-stryker personnel (for example, the p3 maintenance company employed by the hospital).The light has been replaced at the site by a stryker field service technician.
|
|
Event Description
|
It was reported that the light allegedly fell due to the spring arm and horizontal arm separating.There was no injury or adverse consequence reported.
|
|