Hitachi investigated the event and duplicated the error condition in the factory.Due to system controller software anomaly, the couch of verotm/mhi-tm2000 may not move and patient position may not be corrected if all conditions described below are met: (1) calculate the patient positional deviation with exactrac, and press the [c-aut] button to move the couch in order to correct the patient positional deviation.(2) in order to acquire image, press the [x-ray] button within 1 second, after moving the couch.(this action is the likely reason that the couch did not move as directed.) (3) even if the patient positional deviation exceed the allowance value in the image registration of exactrac , the [c-aut] button do not blink (because the couch do not move, the patient positional is not corrected.) (4) press [skip] button to start treatment without verification.(operation that deviates from the normal instructions for use).
|
One of our customers in (b)(6) reported an adverse event to the local distributor on (b)(6) 2017.The distributor notified hitachi on (b)(6) 2017.According to the explanation provided by the customer, we understand that following happened: 1.Patient was positioned with exactrac stereo x-ray normally.2.The therapist pressed the [x-ray] button to verify the position of the patient.3.The therapist tried to move the couch because there was a deviation from the target position, but the couch did not move.4.The therapist did not notice that couch did not move and that the ptv(planning target volume) was not in beam isocenter.The therapist proceeded to apply treatment.5.After the therapists applied several treatment beams they noticed that exactrac showed a deviation of the patient position.6.The therapist performed positioning again and continued the remaining treatment based on the therapist's medical judgment.
|