The photo provided by the customer has been analysed.The photo shows a dilator loaded inside a blue introducer.The distal rim of the introducer showed damage.The introducer material was pushed proximal and showed a possible tear.The venaseal blue introducer, dilator, and j-wire were received.No ancillary devices/components were included.The dilator was inserted inside the blue introducer.The j-wire was inspected and a bend was noted 36 cm from the looped tip.No other damages were noted to the wire.A portion of the distal rim of the introducer was damaged.The rim showed a segment where the material tore and buckled proximally.The diameter of the tip with the introducer inserted was 0.0975¿.This diameter of the distal tip of the dilator with the observed damage exceeds 7f specifications.Damage to the distal tip of the catheter could be seen.If information is provided in the future, a supplemental report will be issued.
|