Catalog Number H938741
Device Problem Misassembled (1398)
Patient Problem No Patient Involvement (2645)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the additive port of an exactamix 3000ml eva tpn bag was not sealed off properly.The customer stated the opening was outside of the bag rather than inside.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
One sample was received for evaluation.Visual inspection and microscopic inspection revealed that the fill port tubing end was detached from the inside of the bag and protruding out in front of the bag.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7064501
MDR Text Key93224229
Report Number1416980-2017-09386
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2020
Device Catalogue NumberH938741
Device Lot Number60082899
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-