Catalog Number H938741 |
Device Problem
Misassembled (1398)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/08/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the additive port of an exactamix 3000ml eva tpn bag was not sealed off properly.The customer stated the opening was outside of the bag rather than inside.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
One sample was received for evaluation.Visual inspection and microscopic inspection revealed that the fill port tubing end was detached from the inside of the bag and protruding out in front of the bag.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|