BOSTON SCIENTIFIC - COSTA RICA (COYOL) GENESYS HTA PROCERVA®; DEVICE, THERMAL ABLATION, ENDOMETRIAL
|
Back to Search Results |
|
Model Number M006580211 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
Burn(s) (1757)
|
Event Date 11/06/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation on (b)(6) 2017 that a genesys hydrothermablation procerva procedure set was used in a hydrothermablation (hta) procedure performed on (b)(6) 2017.According to the complainant, a leakage occurred seven minutes into the ablation phase.Some hot fluid went into the vagina and the control unit displayed a "fluid loss detected" error message.The physician added a forcep into the cervix, continued with the ablation phase and completed the procedure.The patient's condition at the conclusion of the procedure was reported to be "fine".During patient's second visit, she complained on having a vaginal burn.The burn was then treated with a topic cream.The patient was seen by the physician on an unspecified date.The burn was not extensive and was treated with a cream topic.The patient's current condition was good.
|
|