Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (unable to properly communicate, "add gel" messages) was confirmed.Upon investigation the cable connecting the distribution node (dn) to the rear therapy electrodes was pulled from the strain relief, damaging wires in the cable.Additionally, the trunk cable connector latch was cracked, preventing the belt from properly connector the monitor.The root cause for the strained cable was excessive force.The root cause of the broken connector latch was physical abuse.No adverse event resulted from the defective electrode belt.
|