During an inspection on the upper digestive tract, the subject device was used.When the user tried to cut the detainment snare with the subject device, the snare was caught in the cutter part of the device and the user could not cut it.Since the scope could not be removed, the user inserted additional scope and cut the snare using another loop cutter(fs-5q-1).After that, the procedure was completed.No patient injury was reported.
|
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.No object was caught in the cutter part of the device.There were no irregularities with the device.Since the device was manufactured twenty-two years ago, there was no manufacturing record.However, omsc has only shipped devices which passed the inspection.It is assumed that the event occurred because the doctor pulled the slider without positioning the loop vertically to the loop hunger.Although omsc could not find the instruction manual since it is an old product, a similar event occurred to a current similar device, and the instruction manual of the device has warned as follows.-do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible, or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.
|