|
Catalog Number SP-101 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation of blue introducer: the blue introducer was received for evaluation.A kink to the introducer was noted approximately 14.5cm from the distal tip.A dilator from the lab was successfully loaded through the proximal segment; however the dilator could not be loaded past the lodged obstruction.The obstruction was pulled and removed from the introducer.It was discovered the stuck obstruction was the distal end of a dilator.The length of the dilator is approximately 5cm.The fracture face was straight across, consistent with the component being cut.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The physician intended to use vena seal to treat a venous deficiency in the gsv.The device was prepped as per the ifu with no issues identified.No visible damage was noted to the device.The introducer was prepared and positioned in the patient and the dilator was removed.The glue delivery catheter was introduced and approximately halfway into the introducer, the physician experienced resistance.The surgeon tried to continue to advance but the delivery catheter would not pass through the introducer.No issue were noted when dilator had been used.The delivery catheter was removed.Surgeon checked to see if inadvertent glue deployment had occurred but the glue was still approximately 1 cm from the tip.No kinks were noted in the introducer or the catheter.The surgeon opened a second kit and replaced the introducer.The procedure was finished without further issue, one segment was treated and the vein closed.No patient injury was reported.
|
|
Search Alerts/Recalls
|
|
|