Results - bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for broken blades with the incident lot was not observed.Additionally, retention samples were selected for evaluation, and the customer's indicated failure mode for broken blades was not observed as all samples met specifications.Needle points from the sample were visually inspected at x10.No defects were observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion - bd was not able to duplicate or confirm the customers indicated failure mode.
|