AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125225-15
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 11/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported cracked hub and reported leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during preparation, the inflation device was over torqued during connection to the hub resulting in the reported cracked hub; thus resulting in the reported leak.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that a 2.25x15mm xience alpine stent delivery system (sds) was prepared per the instructions for use; however, when it was attached to the indeflator it was noticed that the hub was cracked and it leaked.The sds was not used and the procedure was successfully completed with a new 2.25x15mm xience alpine sds.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7076952
MDR Text Key94377932
Report Number2024168-2017-09340
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648199073
UDI-Public08717648199073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/29/2020
Device Catalogue Number1125225-15
Device Lot Number7081641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial